KI Discus Testing for Safety Cabinets

KI Discus containment testing of microbiological safety cabinets to BS EN 12469

What is KI Discus Testing?

KI Discus testing is a recognised method for assessing the containment performance of Class I and Class II microbiological safety cabinets (MSCs). It uses an aerosolised potassium iodide solution to simulate airborne contaminants and a laser particle counter to measure escape levels outside the cabinet.

This test is required by BS EN 12469 and is essential for laboratories handling biological agents or hazardous particulates where high containment standards are needed.

Why KI Discus Containment Testing Matters

Unlike fume cupboards, safety cabinets are designed to protect both the user and the product within the cabinet from biological and particulate contamination. A failed KI Discus containment test could indicate filter failure or blockage, leakage or recirculation issues, putting staff, samples, and surroundings at risk.

Regular KI Discus testing verifies the integrity of airflow systems, HEPA filters, and fan performance — confirming the cabinet maintains its intended level of protection.

Ki Discuss Testing Process

• The MSC and environment are prepared to standard test conditions
• A fine mist of potassium iodide droplets, produced by a spinning disk, is used as a challenge aerosol to measure the containment of the MSC
• Centripetal collectors sample the air outside the MSC
• The collectors deposit any potassium iodide particles present in the sampled air onto filter membranes
• At the end of the sampling period, the membranes are placed into a palladium chloride solution, where potassium iodide reacts to form visible brown spots
• The spots are carefully counted by eye and compared against the number of droplets produced by the spray generator, allowing the protection factor of the MSC to be calculated
• All open-fronted microbiological safety cabinets are assessed based on this protection factor
• Results are compared against BS EN 12469 thresholds and a detailed certificate is issued

 

Safety Cabinets We Test

• Class I microbiological safety cabinets
• Class II microbiological safety cabinets (recirculating and ducted)
• Cytotoxic isolators and cabinet enclosures
• Cabinets used in pharmaceutical, veterinary, NHS and academic labs

Book Your KI Discus Testing
Book a KI Discus containment test or enquire about safety cabinet servicing

Call 01992 589115